The DODO AQUA pH (DOT-PHS-200) is a submersible industrial online pH sensor measuring 0–14 pH with 0.01 pH resolution and ±0.1 pH accuracy. The signal conditioning is built into the probe, so it outputs a calibrated digital pH value over RS-485 Modbus-RTU with no separate transmitter needed. A pressurised long-life reference rated for over 20 months and automatic NTC temperature compensation keep readings stable. It runs on DC 12–24 V, is IP68 rated with a 3/4" NPT thread, and comes with a 5 m cable and GX12 connector for direct use with the DODO LOGGER or any Modbus master.
حساس DODO AQUA pH (DOT-PHS-200) هو حساس pH صناعي أونلاين قابل للغمر يقيس ضمن مدى 0–14 pH بدقة قراءة 0.01 pH ودقة قياس ±0.1 pH. دائرة معالجة الإشارة مدمجة داخل جسم الحساس، لذلك يُخرج قيمة pH رقمية معايرة عبر RS-485 Modbus-RTU دون الحاجة إلى محوّل منفصل. يتميز بنظام مرجعي مضغوط طويل العمر يدوم أكثر من 20 شهراً مع تعويض تلقائي لدرجة الحرارة بحساس NTC لقراءات مستقرة. يعمل بجهد 12–24 فولت DC، بحماية IP68 وسن تركيب 3/4" NPT، ويأتي مع كابل 5 متر وموصل GX12 للتوصيل المباشر مع DODO LOGGER أو أي جهاز Modbus.
موصل GX12 للتوصيل المباشر مع DODO LOGGER لعرض القراءات وتخزينها ومراقبتها من خلال منصة DODOTIK PLATFORM وتطبيق DODOLINK.
The DODO AQUA pH is part of the DODO AQUA family of industrial online water sensors from DODOTiK. It is a submersible pH sensor with the signal conditioning built into the probe body, so what leaves the sensor is already a calibrated digital pH value on RS-485 Modbus-RTU.
This matters more for pH than for any other measurement. A raw glass electrode produces only a few tens of millivolts at very high impedance, and the most common field failure is the long analog cable between the electrode and a remote transmitter. By digitising at the probe, the DODO AQUA pH removes that weak point entirely: no separate transmitter, no analog pH loop, no high-impedance cable to protect.
The sensing element uses a dual high-impedance differential amplifier feeding a patented reference system. The internal reference solution is held under at least 100 kPa (1 bar) and bleeds outward through a microporous salt bridge, keeping the junction clear of process liquid. This forward bleed is rated for more than 20 months, roughly twice the service life of a conventional industrial reference.
A built-in NTC thermistor applies automatic temperature compensation, so the reported pH is already corrected. The sensor is factory calibrated, and two-point recalibration can be done over the bus using the included pH 6.86, 4.00 and 9.18 buffer powders.
As a standard Modbus-RTU device (default address 6, 9600 8N1), it connects to any PLC, DCS, SCADA or RS-485 master. With the DODO LOGGER, it plugs in directly via the fitted GX12 4-pin connector, logs to microSD, and publishes to the DODOTiK Platform and DODOLINK mobile app with no programming.
Installation note: the probe must be mounted at 15° or more above horizontal, glass end down, and stored wet in 3.5 mol/L KCl solution when out of service.
Specifications
| Parameter | Value |
|---|---|
| Model | DOT-PHS-200 |
| Order code | DA-PH |
| Measured parameters | pH, with temperature as compensation channel |
| pH range | 0–14 pH |
| pH resolution | 0.01 pH |
| pH accuracy | ±0.1 pH |
| Temperature resolution | 0.1 °C |
| Temperature compensation | Automatic (NTC) |
| Reference system | Pressurised microporous salt bridge, 20+ months life |
| Calibration | Two-point: zero at pH 6.86, slope at pH 4.00 or 9.18 |
| Supply voltage | DC 12–24 V ±10 % |
| Power consumption | < 0.3 W @ 12 V |
| Output | RS-485, Modbus-RTU |
| Baud rate | 9600 (default) |
| Frame format | 8N1 |
| Default slave address | 6 (changeable, 1–127) |
| Body diameter | Φ30 mm |
| Electrode guard diameter | Φ22 mm |
| Overall length | 185 mm |
| Wrench flat | 28 mm |
| Wetted material | POM |
| Mounting | Immersion, 3/4" NPT thread |
| Mounting angle | 15° or more above horizontal, glass end down |
| Working pressure | < 0.2 MPa |
| Working temperature | 0 to +65 °C |
| Ingress protection | IP68 |
| Cable | 5 m, Φ6 mm shielded, corrosion-resistant jacket |
| Connector | GX12 4-pin female (VCC, GND, A, B) |
| In the box | Sensor, 3 calibration powder sachets (6.86 / 4.00 / 9.18), test certificate |
| Quality testing | 72-hour burn-in on every unit |
Applications