The DODO AQUA CL (DOT-FCS-200) is an industrial amperometric free chlorine sensor that measures hypochlorous acid (HClO) directly across 0–2.000 mg/L with 0.001 mg/L resolution and a response time under 30 seconds. Using the constant-voltage method with automatic Pt1000 temperature compensation, it delivers the compliance-grade reading that ORP only approximates. It is supplied with its own flow cell for a controlled 30–60 L/h sample stream, outputs over RS-485 Modbus-RTU, runs on DC 12–24 V at just 0.2 W, and ships with a 5 m cable and GX12 connector for plug-in use with the DODO LOGGER.
حساس DODO AQUA CL (DOT-FCS-200) هو حساس كلور حر أمبيرومتري صناعي يقيس حمض الهيبوكلوروز (HClO) مباشرة ضمن مدى 0–2.000 ملغم/لتر بدقة قراءة 0.001 ملغم/لتر وزمن استجابة أقل من 30 ثانية. يعتمد على طريقة الجهد الثابت مع تعويض تلقائي لدرجة الحرارة بحساس Pt1000، ليقدم قراءة مطابقة لمتطلبات الامتثال بدلاً من التقدير الذي يوفره ORP. يأتي مع خلية تدفق خاصة به لتيار عينة مضبوط بمعدل 30–60 لتر/ساعة، ويعمل بواجهة RS-485 Modbus-RTU وبجهد 12–24 فولت DC باستهلاك 0.2 واط فقط، مع كابل 5 متر وموصل GX12 للتوصيل المباشر مع DODO LOGGER.
موصل GX12 للتوصيل المباشر مع DODO LOGGER لعرض القراءات وتخزينها ومراقبتها من خلال منصة DODOTIK PLATFORM وتطبيق DODOLINK.
The DODO AQUA CL is part of the DODO AQUA family of industrial online water sensors from DODOTiK. It continuously measures free chlorine, the hypochlorous acid (HClO) that actually does the disinfecting, from 0 to 2.000 mg/L at 0.001 mg/L resolution.
This is the measurement that regulators ask for. ORP tells you that water is oxidising; free chlorine tells you how much disinfectant is actually in it, in the same units the compliance report is written in. Installed at the outlet of a treatment works or at the far end of a distribution network, it turns chlorine dosing from a scheduled job into a controlled process.
Measurement is amperometric, using the constant-voltage (three-electrode) method. A fixed potential is held between the working and reference electrodes; free chlorine diffuses through a PTFE membrane and is reduced at the working electrode, producing a current proportional to its concentration. Because the potential is held constant rather than swept, the reading settles in under 30 seconds. A built-in Pt1000 element provides automatic temperature compensation and reports temperature as its own channel.
Unlike the other DODO AQUA water-quality sensors, this one is not an immersion probe. It is installed in the supplied flow cell with a controlled 30–60 L/h sample flow across the membrane, which keeps the measurement accurate and stable. It is specified for media between pH 4 and pH 9 and 5–50 °C. The sensor is factory calibrated, with optional two-point recalibration: zero in chlorine-free water and slope against a standard.
As a standard Modbus-RTU device (default address 6, 9600 8N1), it connects to any PLC, DCS, SCADA or RS-485 master. With the DODO LOGGER, it plugs in directly via the fitted GX12 4-pin connector, logs to microSD, and publishes to the DODOTiK Platform and DODOLINK mobile app with no programming. It can share the same bus with the DODO AQUA pH for combined monitoring.
Note: a new electrode must be activated in tap water for 24 hours before first use. The electrode is a consumable part with a 6-month warranty.
Specifications
| Parameter | Value |
|---|---|
| Model | DOT-FCS-200 |
| Order code | DA-CL |
| Measured parameters | Free chlorine (HClO) and temperature |
| Measuring principle | Amperometric, constant-voltage method |
| Measuring range | 0–2.000 mg/L |
| Resolution | 0.001 mg/L |
| Accuracy | ±5 % of reading or ±0.05 mg/L |
| Response time | < 30 s |
| Temperature range | 5–50 °C |
| Temperature accuracy | ±0.5 °C |
| Temperature resolution | 0.1 °C |
| Medium pH range | pH 4–9 |
| Sample flow rate | 30–60 L/h |
| Temperature compensation | Automatic (Pt1000) |
| Calibration | Two-point: zero in chlorine-free water, slope against standard |
| Supply voltage | DC 12–24 V ±10 % |
| Power consumption | 0.2 W @ 12 V |
| Output | RS-485, Modbus-RTU |
| Baud rate | 9600 (default) |
| Frame format | 8N1 |
| Default slave address | 6 (changeable, 1–127) |
| Body diameter | Φ30 mm |
| Electrode guard diameter | Φ22 mm |
| Overall length | 185 mm |
| Wrench flat | 28 mm |
| Wetted material | POM and PTFE |
| Mounting | Flow cell (supplied), not an immersion probe |
| Working pressure | < 0.1 MPa |
| Ingress protection | IP68 |
| Cable | 5 m, 4-core shielded twisted pair |
| Connector | GX12 4-pin female (VCC, GND, A, B) |
| In the box | Sensor with cable, flow cell, test certificate |
| Warranty | 1 year on sensor body, 6 months on electrode |